Chapter 21 Design of implants and factors associated with implant failure (wear, loosening)

📄 pp. 1278–1343 (PDF)Book: Postgraduate Orthopaedics Viva Guide

📝 Reduction summary (~35% shorter) — high-yield viva edition. All figures & classifications retained; key points bolded for speed reading. Source page badges preserved. Full text: postgradviva2026.pages.dev
source p. 1278

Tribology of natural and artificial joints

Kiran Singiseƫ and Paul A. Banaszkiewicz

source p. 1279

Introduction#

A classic example is the clinical photograph of an explanted worn PE cup leading on to a discussion of wear.

A good understanding of tribological properft es helps the orthopaedic.

Wear is an A-list topic with similar competency questions in the first part of a viva but unexpected or esoteric higher-order thinking questions in the second part.

source p. 1280

Structured oral examination question 1#

source p. 1281

Clinical photograph of explanted poly cup with wear and cement

EXAMINER
What do you see? Why has this happened?
CANDIDATE
This is a clinical picture of an explanted PE cup demonstrating acetabular wear (Figure 21.1).
Figure 21.1
Figure 21.1Figure 21.1 Clinical photograph of explanted polyethylene cup demonstrating worn surface.p. 1285
EXAMINER
There is quite obvious wear seen on the inside of the acetabular cup, so what do you think has been the most likely cause for revision?
CANDIDATE
Aseptic loosening.
EXAMINER
What do you mean by aseptic loosening?
CANDIDATE
Aseptic (i.e. not caused by infection) loosening refers to the failure of fixation a t the bone/implant interface.
EXAMINER
What is the difference between aseptic loosening and osteolysis.
CANDIDATE
Aseptic loosening is an umbrella term that is used to describe total joint arthroplasty failure resulting from inadequate initial fixation, mechanical loss of fixation overtime, or biological loss of fixation caused by particle-induced osteolysis around the implant.
EXAMINER
What do we mean by wear?
CANDIDATE
Wear is the removal of material from two surfaces underload, due to a sliding motion between them. Learn one that you are happy with and stick to it.
EXAMINER
What are the different types of wear that can occur?
CANDIDATE
There are two broad categories of wear, (1) mechanical and (2) chemical. Mechanical involves: Abrasive. Adhesive. Fatigue.
source p. 1282

Erosive.

Chemical is independent of load and sliding distance:

Corrosion.

EXAMINER
What are the various types of mechanical wear?
CANDIDATE
Asperities on the hard bearing carve ridges (plough/cheese grater effect1) into the softer bearing. An example is between a metal femoral head and a polyethylene liner.
EXAMINER
What do you mean by adhesive wear?
CANDIDATE
Adhesive wear occurs when opposing asperities of two surfaces bond with each other to form a junction .2 This junction is held by intermolecular bonds and generates friction.
EXAMINER
What is fatigue w ear?
CANDIDATE
Cracks propagate, joining together, and the loose material comes away from the surface. An example is delamination of the polyethylene inT KAs.
EXAMINER
What is erosive wear?
CANDIDATE
Erosive wear occurs when hard particles travelling in fluid interposed between two surfaces remove some of the surface as they collide into it.
EXAMINER
Corrosive wear?
CANDIDATE
This is inviting tr ouble.3 Concentrate on the three big wear mechanisms (abrasive, adhesive and fatigue w ear) as these are what the examiners are most familiar with.
source p. 1283

Although erosive, the examiners will decide if they want to further probe you for extra details.4.

Third-body wear can be either classified as a subtype of abrasive wear or as a separate, distinct type of wear mechanism.

Erosive wear is classified as either a major or minor cause of mechanical wear depending on which textbook is read.

EXAMINER
Can you drawout the different types of wear (Figure 21.2a–21.2c)?
Figure 21.2a
Figure 21.2aFigure 21.2a Wear mechanisms. Adhesive wear: opposing asperities bond to each other and shear off as one surface slides over the op. 1285
COMMENT
In the exam it is difficult to draw if you are caught cold and need to work it out for the first timeT o be slick, the drawings need to be practised beforehand.
EXAMINER
What type of wear has occurred on the PE cup surface?
CANDIDATE
The majority of the wear pattern would be abrasive. Third-body abrasive wear is also a possibility. This is like having sand in one’s shoes.5 It is a form of abrasive wear that occurs when a hard particle becomes embedded in a soft surface.
EXAMINER
How is this wear mechanism in the hip different to that of the knee?
CANDIDATE
Adhesive and abrasive wear is more pronounced for THA while fatigue w ear (piting and delamination) is more problematic for TKA.
EXAMINER
What is the RANKL pathway?
COMMENT
Candidates should be familiar with the RANKL pathway from the Part 1 SBI/EMI paper, but there is a world of difference in answering this topic in a viva exam. The main biological system that leads to osteolysis-induced resorption of bone is the receptor activator of nuclear factor-κB (RANK)/RANK ligand (RANKL) axis Activation of this system results in enhanced osteoclast recruitment and activity adjacent to bone implant surfaces, leading to osteolysis.

Once macrophages are activated by particulate debris, the bone-resorbing cells directly responsible for the pathogenic bone loss in osteolysis.

Other cell types involved in the production of cy tokines and inflammatory mediators include osteoblasts and fibroblasts.

Research suggests that UHMWPE increases the release of RANKL from osteoblasts, while OP Gis significantly inhibited.

Although wear debris may consist of polyethylene, PMMA cement, or metal, by far the great majority of wear particles derives from polyethylene.

EXAMINER
What factors affect the degree of osteolysis from wear particles?
CANDIDATE
Factors affecting osteolysis severity include: 1. Particles within the br oad size range of 0.1–10.0 μm are phagocytosed by macrophages, leading to cellular activation. Those in the size range 0.1–1.0 μm are the most active. 2. Shape of particles (elongated particles are more active compared to round or spherical particles). 3. Volume of particles (the critical volume is 140 mm3/year). 6. Immune response to particles.
EXAMINER
What areG ruen zones?
CANDIDATE
This is a widely used system in which the femoral component interface is considered in seven zones (Figure 21.4).
Figure 21.4
Figure 21.4Figure 21.4 Gruen zones. Remember 1–7 starting a tGT and ending at calcar.p. 1288
source p. 1285

It is the progressive changes seen in serial radiographs that are important in diagnosing osteolysis and femoral stem loosening.

Figure
Figurep. 1285

Figure 21.1 Clinical photograph of explanted polyethylene cup demonstrating worn surface.

Figure
Figurep. 1285

Figure 21.2a Wear mechanisms. Adhesive wear: opposing asperities bond to each other and shear off as one surface slides over the other.

source p. 1286
Figure
Figurep. 1286

Figure 21.2b Wear mechanisms. Abrasive wear: asperities on the harder material cut into the asperities of the softer material. The new particles become third bodies.

Figure
Figurep. 1286

Figure 21.2c Wear mechanisms. Fatigue w ear: cyclical loading causes accumulation of micr o-damage that breaks off as wear particles.

source p. 1287
Figure
Figurep. 1287

Figure 21.3 Model of interplay between macrophages, fibroblasts, lymphocyte, osteoclasts and osteoblasts in periprosthetic osteolysis. Particles may stimulate macrophages, fibroblasts and osteoblasts directly to induce RANK Land pro-inflammatory cytokines that can induce RANK LItis thought that T cells stimulated by the pro-inflammatory microenvironment may also promote osteoclast formations ynergized with TNF-α, by secreting IL-17. Thus, RANKL , TNF-α, IL-1, IL-6, IL-17 and M-CSF may mediate the differentiation of myeloid precursor cells into multinucleated osteoclasts.6

Experimentally, polymethylmethacrylate (PMMA) and polyethylene (PE) particles have been shown to activate macrophages via the TRL pathway.

source p. 1288
Figure
Figurep. 1288

Figure 21.4 Gruen zones. Remember 1–7 starting a tGT and ending at calcar.

source p. 1289

Structured oral examination question 2#

source p. 1290

Wear and osteolysis

A clinical picture of a worn polyethylene cup is shown (Figure 21.5).

Figure 21.5
Figure 21.5Figure 21.5 Clinical picture demonstrating worn surface of retrieved PE cup.p. 1292

Mechanisms of wear.

Osteolysis.

Effective joint space.

EXAMINER
What are the mechanisms of wear?
CANDIDATE
The main mechanisms of wear include abrasive, adhesive and fatigue w ear (see previous answer).
EXAMINER
What do we mean by the term ‘osteolysis’?
CANDIDATE
The core of the biological response that leads to osteolysis involves the receptor activator of NF-κB ligand [RANKL]–RANK axis for osteoclast precursors, resulting in their differentiation and maturation (see previous answer).
EXAMINER
What about the importance of macrophages in the pathogenesis of osteolysis?
CANDIDATE
The cellular response that occurs in osteolysis is dominated by macrophages. Particles ranging from 0.1 to 10 μm in diameter undergo phagocytosis by macrophages. Once activated by particulate debris, macrophages secrete various kinds of mediators to incite a complex cascade of events culminating in osteoclast maturation. Pro-inflammatory mediators such as PGE2, TNF-α and IL-6 are generated in abundance by particle- challenged macrophages.
EXAMINER
What do we mean by effective joint space7?
CANDIDATE
2. All periprosthetic regions that are accessible to joint fluid and its particulate debris. 3. The effective joint space is a concept that describes the entire volumetric area within a hip joint construct that can be infiltrated by PE wear particles and macrophages.
COMMENT
The presence of particulate matter in joint fluid will initiate a localized macrophage-induced phagocytosis and result in bone resorption. If joint fluid is distributed more evenly in an interface, there will be slower resorption of bone accompanied by a fibroblastic response resulting in the radiographic appearance of linear (diffuse) bone loss.
EXAMINER
How can you reduce effective joint space?
CANDIDATE
Reduction in the effective joint space may reduce the amount of osteolysis that can occur. The use of bone screws for fixation of acetabular shells is thought to create new voids in the acetabular bone that increase the effective joint space. Implanting acetabular shells without any screw holes in theory reduces the effective joint space.
EXAMINER
What new designs of hip replacements have been introduced to retard osteolysis by limiting the generation and spread of particulate debris?
CANDIDATE
Improved liner locking mechanisms reduce the amount of motion between shell and PE liner. It has been suggested wear debris can be produced at the interface between the metal acetabular shell and PE liner. Candidates could be moved easily towards discussing improvements in PE manufacture, sterilization, shelf packaging, annealing versus heating , amorphous versus crystalline phase, etc. as part of an evolving viva on wear (advanced questions).
EXAMINER
What are the risk factors for osteolysis?
CANDIDATE
Risk factors can be broken down into patien t-, surgical- and prosthesis-related factors. Implant factors include prosthetic design, bearing couple, PE (manufacturing process, post- manufacturing sterilization, thickness of PE insert (knee) and liner (hip)).
Figure
Figurep. 1292

Figure 21.5 Clinical picture demonstrating worn surface of retrieved PE cup.

source p. 1293

Structured oral examination question 3#

source p. 1294

Wear in THA

Picture of aseptic loosening , what is wear, measures to reduce wear, particle size.

A radiograph of aseptic loosening hip is shown to the candidate (Figure 21.6).

Figure 21.6
Figure 21.6Figure 21.6 Anteroposterior (AP) radiograph, pelvis, demonstrating loose right THA. Cement fracture and femoral stem and cup migrap. 1296
EXAMINER
What is wear?
CANDIDATE
Mechanical wear is the removal of material from two surfaces underload, due to the sliding motion between them.
EXAMINER
What are the modes of wear of artificial joints?
CANDIDATE
The four modes of wear are: Mode 1. Mode 2. Mode 3. Mode 4. McKellop’s classification. Do not confuse with the four Gruen modes of failure of cemented femoral stems or vice versa.9 The fundamental mechanisms of wear include adhesive, abrasive and fatigue w ear.
EXAMINER
What measures can betaken to reduce wear?
COMMENT
There is more than enough material to discuss that would use up the full 5 minutes of viva discussion if you are allowed to keep on talking (unlikely, but worth trying). So, a candidate’s leadin phrasing to discuss PE wear could be ‘The main type of wear particle implicated in osteolysis and loosening of total joint replacements is polyethylene. Methods to improve the wear characteristics of PE include ’: Manufacturing techniques.
source p. 1295

Another possible option ist o discuss:

Surgeon (technique) factors:

Implant selection, a voidance of implant malalignment, accurate restoration of mechanical axis joint.

Patient factors:

Weight (weight reduction).

Activity le vel (avoidance of excessive activities, eg. waterskiing, treadmill running, etc.).

Implant design.10 Decide on whether to discuss hips or knees or both.

Hips.

Offset. Decreasing offset increases joint reaction forces.

Choice of bearing couple (MoP, CoP, MoM).

Head size.

Knees.

Conformity.

Thickness of PE (minimum 8 mm).

Femoral rollback.

source p. 1296
Figure
Figurep. 1296

Figure 21.6 Anteroposterior (AP) radiograph, pelvis, demonstrating loose right THA. Cement fracture and femoral stem and cup migration. Gruen mode 1a failure.

source p. 1297

Structured oral examination question 4#

source p. 1298

Wear in TKA

Polyethylene wear in TKA: discuss all factors.11

EXAMINER
What does the picture show (Figure 21.7)? CANDIDATE 1: A worn tibial tray. CANDIDATE 2: A worn PE tibial insert. It looks like it is a CR-retaining ploy. There is evidence of severe delamination with uneven wear more pronounced medially. Discoloration is as a dark yellow tint representative of polyethylene oxidation. Discuss the factors associated with PE wear in TKA.
Figure 21.7
Figure 21.7Figure 21.7 Worn PE tibial tray. Delamination is seen as thin sheets of polyethylene separated from the surface.p. 1300
CANDIDATE
Ultra-high molecular polyethylene wear debris triggering osteolysis is one of the major causes of failure of TKA. Varus alignment of implants leads to accelerated medial PE wear and the risk of early failure. Varus placement of the tibial component > 3° leads to almost double the PE volumetric penetration rate. This is compounded by the roughness of the metal tibial tray, resulting in increased PE particle generation leading to osteolysis. This can be particularly severe in tibial tr ays with holes for fixation screws around which osteolytic lesions can develop. Additional component features that increase wear include thinner polyethylene inserts, some non- cemented tibial baseplates supplemented with tibial screws, and metal-backed patellar components. The change from sterilization of PE by gamma radiation in airt o an inert gas has resulted in much lower rates of osteolysis at 10 years post surgery.
source p. 1299

The mechanism of wear between these two joints is different.THA wear is mostly due to micro-adhesion.

Most supporting evidence for HXPLE in knee arthroplasty is derived from in-vitro wear simulator studies that show a reduction of w ear of up to 60%.

conventional UHMWPE bearings have all found no significant difference in clinical or radiological outcomes between the two bearings.12,13,14.

HXLPE had a lower cumulative percentage revision than conventional polyethylene at 5 years (4.0% vs 2.6%) and

10 years (5.8% vs 3.6%).

It is recognized that failed TKA have larger flake-shaped debris, which elicits a tissue response characterized by fewer macrophages.

COMMENT
A clinical picture of a tibial insert demonstrating a white subsurface oxidized band of PE is another classic lead-in prop to discuss PE wear in TKA.
EXAMINER
What are the wear mechanisms in TKA?
CANDIDATE
Three main wear mechanisms can beseen in TKA.15 These are adhesive, abrasive and surface fatigue. Adhesive wear: the bonds formed between different materials are stronger than the specific material properties of either surface and therefore pullout fragments from one surface to another. Abrasive wear: a harder rougher material ploughs through a softer material.
source p. 1300
Figure
Figurep. 1300

Figure 21.7 Worn PE tibial tray. Delamination is seen as thin sheets of polyethylene separated from the surface.

Figure
Figurep. 1300

Figure 21.8 Retrieval PE insert demonstrating classic white band defect of oxidation located 1–2 mm below the machined surface of PE.

EXAMINER
What wear damage occurs at the tibial PE surfaces of aTKA?
CANDIDATE
Hood et al.16 described seven types of wear mechanism damage at the articulating surfaces of TKA: Burnishing (polishing).
source p. 1301

Burnishing: contact areas are polished due to a combination of abrasive and adhesive wear.

Scratching: this is caused by abrasive wear.

Abrasion: characterized as a shredding of the polyethylene surface and classified as a mode of abrasive wear.

Pit inga mode of fatigue w ear that is characterized by the formation of millime tre-sized craters

(Figure 21.9). It is considered to be a more benign wear mechanism that does not provoke an osteolytic response.

Figure 21.9
Figure 21.9Figure 21.9 Piting small cr ater-like surface defects.p. 1301

Third-body wear: wear debris can act as third-body particles, initiating w ear by rubbing at the bearing surfaces.

Delamination ( Figure 21.10), and can result in catastrophic wear.

Figure 21.10
Figure 21.10Figure 21.10 Delamination of PE.p. 1302

There is gross disruption of the material to a depth of 0.5 mm or more due to the formation and propagation of subsurface cracks.

Figure
Figurep. 1301

Figure 21.9 Piting small cr ater-like surface defects.

source p. 1302
Figure
Figurep. 1302

Figure 21.10 Delamination of PE.

More severe plastic deformation of the tibial insert may be an indication of malalignment or a mismatch of component sizes.

PE wear in knee arthroplasty occurs from a combination of rolling, rotation motions between the bearing surfaces.

Strategies to reduce polyethylene wear include the following.

Improving implant design.

Congruent bearing designs lower the amount of cross-shear stresses.

The sagift al plane should be concave or dished and the individual medial and lateral tibial plateaus should also be dished in the coronal plane.

Improved locking mechanisms of modular tibial components to reduce potential back side wear.

Highly polished tibial baseplate.

A rotating yet flat PE bearing is matched against a highly polished cobalt chromium surface.

Monobloc tibial components. The PE bearing surface is direct compression-moulded to the tibial baseplate.

All polyethylene tibial components have been used in an attempt to decrease or eliminate the problems associated with backside wear.

source p. 1303

Improvements in the quality of ultrahigh-molecular-weight PE.

Improved sterilization techniques (gamma irradiation inert atmosphere).

Development of newer, highly cross-linked PE with the introduction of vitaminE and sequential annealing.

Refining surgical techniques.

Computer navigation.

Fellowship-trained surgeons.

EXAMINER
Is there any evidence that all-polyethylene tibial components reduce wear (Figure 21.11)?
Figure 21.11
Figure 21.11Figure 21.11 All-polyethylene tibial component (APT), posterior stabilized design.p. 1305
CANDIDATE
With all-polyethylene tibial (AP T; Figure 21.11) the same amount of tibial resection allows for a thicker PE to be used, potentially increasing the lifetime of the prosthesis if wear rates are equivalent, as less tibial resection will be required to achieve the same poly thickness as a design with a metal tray, resulting in a larger metaphyseal surface area and the ability to use larger tibial component sizes, reducing the magnitude of contact stresses transmift ed across the joint while also preserving metaphyseal bone stock.17
Figure 21.11
Figure 21.11Figure 21.11 All-polyethylene tibial component (APT), posterior stabilized design.p. 1305
EXAMINER
What do we mean by conformity?
CANDIDATE
Contact stresses experienced at the PE surface are inversely proportional to the degree of conformity between the femoral condyle and the tibial PE insert.
EXAMINER
So why don’t we just go with highly conforming knee designs?
CANDIDATE
Highly conforming TKA designs significantly increase the stresses transmift ed to the fixation interface and increase the risk of early aseptic loosening.
EXAMINER
What about mobile bearings?
source p. 1304
CANDIDATE
In-vitro wear studies have shown that mobile bearing produces less wear compared to fixed designs.
EXAMINER
What about surgeon factors?
CANDIDATE
Surgeon-controlled strategies recommended for reduction of PE w ear include meticulous attention to ligament balancing.
EXAMINER
What else?
CANDIDATE
Ideally, a fellowship-trained arthroplasty surgeon.
EXAMINER
What else?
CANDIDATE
GIRFT (geting it right first time). Computer navigation.
EXAMINER
Is there any evidence that computer navigation improves the accuracy of implant positioning?
CANDIDATE
There is some evidence that computer navigation reduces the number ofT KAs that have a coronal malalignment of more than 3°.
EXAMINER
Which paper?
CANDIDATE
I am not familiar with the specific papers, but the general literature suggests better alignment with navigation.
source p. 1305
Figure
Figurep. 1305

Figure 21.11 All-polyethylene tibial component (APT), posterior stabilized design.

source p. 1306

Reference#

Bauwens K, Matihes G, Wich M, et al. Navigated total knee replacement. A meta-analysis. J Bone Joint

Surg [Am]. 2007;89A:261–269.

CANDIDATE
Although coronal malalignment is reduced, mean alignment and mechanical axis did not differ between navigated and conventional TKRA groups.
EXAMINER
Is there any evidence that computer navigation improves the survival of a knee prosthesis?
CANDIDATE
It is very difficult to prove that computer navigation definitely reduces the need for revision surgery. Recent Australian Registry data suggest there may be a small advantage, particularly in younger patients, as there is a small reduction in the rate of revision for loosening in this group.
EXAMINER
What about the use of alternative bearings?
CANDIDATE
Some surgeons have begun using oxidized zirconium femoral components as a means of reducing polyethylene wear. This technology incorporates a zirconium oxide ceramic coating on a zirconium metal alloy femoral component.
EXAMINER
Any long-term results reported?
CANDIDATE
Ten-year results were reported by Pinczewski et al. from Australia.18 They showed comparable rates of survival with other implants and excellent functional outcomes 10 years postoperatively.
EXAMINER
Any concerns with the paper?
CANDIDATE
Single-surgeon series in a tertiary specialized referral centre that may not reflect the average standard knee arthroplasty surgeon practice.
source p. 1307
Figure
Figurep. 1307

Figure 21.12 Oxium-coated femoral knee implant. During manufacture, OXINIUM implants undergo a process that transforms the implant’s surface into a hard, ceramicized metal.

source p. 1308

Structured oral examination question 5#

The candidate is shown a clinical picture of catastrophic PE failure in TKA (Figure 21.13).

Figure 21.13
Figure 21.13Figure 21.13 Catastrophic failure of PE tibial insert.p. 1308
COMMENT
Remember to structure your answer (1) PE thickness, (2) articular surface design, (3) knee kinematics, (4) PE manufacture, (5) PE sterilization and (6) surgical technique. This is a clinical photograph of an explanted tibial knee replacement component demonstrating catastrophic PE failure.
Figure
Figurep. 1308

Figure 21.13 Catastrophic failure of PE tibial insert.

source p. 1309

Structured oral examination question 6#

source p. 1310

Charnley THA

Polyethylene cup: what is polyethylene, manufacturing advances, wear, what is the stem made from, stress–strain of the stem

EXAMINER
What is polyethylene?
CANDIDATE
UHMWPE is a member of the polyethylene family of polymers with the repeat unit [C2H4]n, with n denoting the degree of polymerization. It is a linear (non-branching), semi-crystalline polymer which can be described as a two-phase composite of crystalline and amorphous phases.
EXAMINER
How is PE manufactured?
CANDIDATE
Polyethylene resin is simultaneously heated and pressurized within an evacuated chamber. As the solid polyethylene forms, it is extruded through an open extrusion port within the chamber. A primitive method of polyethylene manufacture, but cheap. Because the extrusion process is non-continuous, inconsistencies can be found within the solid polyethylene bar stock. Calcium stearate crystals could be found between the particles of polyethylene, resulting infusion defects that became the point of crack initiation and propagation.
source p. 1311

The resin is directly moulded into the finished implant.

There is no secondary machining of the bearing surface. Best wear profile.

Direct compression moulding has a lower susceptibility to fatigue cr ack formation and propagation.

Hot isostatic pressing (HIPing) into bars.

Multistep conversion process of resin powder into stock material.

EXAMINER
What do we mean by the cuting tool effect of polyethylene?
CANDIDATE
In a machined polyethylene insert, machine marks from the lathe create numerous micron- size grooves and shreds on the bearing surface. As the high-speed cuting la the removes PE, the remaining nearby PE is stretched. Stretching occurs in the amorphous areas of PE and is most pronounced in the PE 1–2 mm below the surface of the cut PE.
EXAMINER
What manufacturing advances have occurred with PE?
CANDIDATE
The advent of highly cross-linked PE (HXLPE) has been shown to improve wear rates in hip arthroplasty. Compared to standard PE, HXLPE has: Better wear resistance. To score a 7 a candidate may need to discuss supportive literature of improved implant survivorship with HXLPE in THA. To score an 8, a candidate may need to discuss the controversies of HXLPE use in TKA.
EXAMINER
What do we mean by cross-linking?
source p. 1312
CANDIDATE
By removing atoms or side groups from adjacent chains, covalent bonds can form that link chains together, inhibit relative molecular movement and as such modify the physical properties of the materials.
EXAMINER
What methods are used for PE sterilization?
CANDIDATE
Sterilization methods can be divided into (1) non-energetic (nor adiation) and (2) energetic (using radiation methods. Both ethylene oxide and gas plasma sterilization a void free radical production but do not allow cross-linking of PE. As such, they have a higher wear rate compared to cross-linked PE. Gamma sterilization in air: this makes PE susceptible to oxygenation. Irradiation in the presence of oxygen leads to chain scission of the polyethylene long chain and free radical generation a t the crystal surfaces. Cross-linked PE has improved resistance to adhesive and abrasive wear. This results in an increase in wear resistance. Two different approaches are adopted to achieve oxidation resistance. 1. Annealing: this involves a single thermal treatment below the melting temperature of UHMWPE so that crystallinity and mechanical properties are preserved. 2. Remelting: pos t-irradiation remelting of the polymer above the crystalline transition.
EXAMINER
What is the Charnley femoral stem made of?
source p. 1313
CANDIDATE
The original stem was a monoblock, flat-backed design with a polished surface manufactured out of EN58J stainless steel (Figure 21.14).
Figure 21.14
Figure 21.14Figure 21.14 Radiograph of Charnley total hip arthroplasty.p. 1313
Figure
Figurep. 1313

Figure 21.14 Radiograph of Charnley total hip arthroplasty.

Score 7

Rates of stem fractures led to change in stem surface in 1969 to am aft finish using the vaquasheen process, which deliberately surface-hardened the metal to resist fracture.

source p. 1314

In 1975, anteroposterior Cobra flanges were added to the stem to prevent the escape of cement at the level of the neck resection and also pressurize the cementin the femoral canal.

These changes resulted in a shift in the behaviour of the stem to a composite beam where its predecessor had obeyed the taper-slip principle.

In 1982, the material was again changed to Ortron 90, a cold-worked stainless steel with high fatigue strength.

Gold medal

COMMENT
When Charnley changed his stem from the flatback to the Cobra design, the biomechanical characteristics were changed from a tapered polished (force-closed) to a shape-closed or composite beam biomechanical design. If you are feeling very confident you can discuss the change in stem behaviour with change in Charnley stem.
EXAMINER
What about the stress–strain curve of the stem?
CANDIDATE
The slope of the stress–strain curve of the stem would be the elastic modulus of stainless steel.
EXAMINER
What are the risk factors for femoral stem fracture (Figure 21.15)?
Figure 21.15
Figure 21.15Figure 21.15 Explanted picture of broken Charnley femoral stem.p. 1315
CANDIDATE
Multiple risk factors for prosthetics tem fracture include: High BMI. Varus stem positioning.
EXAMINER
What is the incidence of Charnley stem fracture?
CANDIDATE
The first generation of Charnle y stainless steel femoral stems fractured in approximately 4.1% of patients.
source p. 1315

Improvements in stem design and metallurgy have markedly reduced the incidence of femoral stem fracture.

COMMENT
This is the basic science viva, so candidates are less likely to be asked how they would revise a broken femoral Charnley stem.
Figure
Figurep. 1315

Figure 21.15 Explanted picture of broken Charnley femoral stem.

source p. 1316

Structured oral examination question 7#

source p. 1317

TK Aloose implant

Same discussion of wear, mechanics of loosening and biology of osteolysis as in previous questions.

Discussion about cement – what is it, materials, properties, etc.

Wedges for reconstruction of tibial defect, how wedges work, why wedges and not cement for buildup etc.

EXAMINER
What types of tibial bone loss can occur with primary THA?
CANDIDATE
Large posteromedial asymmetrical osseous defects are often seen in the proximal tibia while performing a primary total knee arthroplasty (TK Ain severe varus knees. Depending upon the size of the defect, these can be treated with cementoplasty, structural bone gratis or metallic wedges.
EXAMINER
What about cement?
CANDIDATE
Cement is cost-effectiv e, but it cannot be used to address large bony defects. Thermal necrosis of bone can occur as well as shrinkage when used in large quanties.
EXAMINER
How do augments work for build-up of bone defects?
CANDIDATE
Metal augments allow rapid filling of bone defects that have been geometrically shaped with instruments. Using thick augments (30 mm) may result in a painful subcutaneous bulge in the tibia or in the distal femur. Therefore, to achieve stable fixation a pr ess-fit metaphyseal femoral sleeve can be used to enhance the rotational stability of the femoral component.
EXAMINER
Anything else that can be used to deal with bone loss?
CANDIDATE
Extensive bone loss can beseen with revision TKA. A broach technique is used to prepare the bone for the press-fit implant.
source p. 1318

Metaphyseal filling titanium cones (Figure 21.17a and 21.17b).

Figure 21.17a
Figure 21.17aFigure 21.17a and 21.17b Metaphyseal filling titanium cones used to reconstruct metaphyseal bone loss in the tibia or femur.p. 1319

I would use impacted morsellized bone graft when the reis a contained bone loss larger than 10 mm. I would avoid its use if there was significant cortical bone loss or uncontained defects.

Megaprostheses: high complication rate but the surgical procedure and rehabilitation are rapid.

High rate of infection (~5%) that can often be followed by amputation.

Figure
Figurep. 1318

Figure 21.16 Metaphyseal sleeve. The broad tibial sleeve contacts the remaining proximal tibial cortex impacting stability.

source p. 1319
Figure
Figurep. 1319

Figure 21.17a and 21.17b Metaphyseal filling titanium cones used to reconstruct metaphyseal bone loss in the tibia or femur.

source p. 1320

Structured oral examination question 8#

source p. 1321

Cementing technique THA

Picture of broken cement mantle (Figure 21.18) – reasons for this.

Figure 21.18
Figure 21.18Figure 21.18 Anteroposterior radiograph showing the Exeter total hip replacement with radiolucent lines around the femoral componep. 1323
EXAMINER
What are the reasons for a broken cement mantle?
CANDIDATE
The early development of stem–cement interface debonding (separation) and subsequent cement fracture are thought to be the initiating events of aseptic loosening. It is important to reduce cement stresses so as to minimize the risk of cement debonding and fracture.
EXAMINER
So how can high cement stresses be avoided?
CANDIDATE
By the creation of an optimally thick symmetric and homogeneous cement mantle.
EXAMINER
So how do we achieve this?
CANDIDATE
Stresses experienced in the cement mantle have been shown to be highest at the stem tip and secondarily at the proximal–medial cement mantle. Stem malalignment produces non-uniform cement mantle thickness in key areas.
EXAMINER
Can you be more specific19?
CANDIDATE
A proximal–medial cement mantle greater than 10 mm orless than 2 mm in thickness is associated with a significant increase in cement fracture, radiolucent lines at the prosthesis–cement interface and progressive component loosening when compared to proximal–medial cement mantles that measure 2–5 mm in thickness.20
source p. 1322

An asymmetrical distal cement mantle significantly increases the risk of implant failure.

Inadequate centralization of the s tem or malrotation will result in excessively thinned areas of distal cement, increased cement strains, prosthesis bone contact.

Preserving < 2 mm of proximal–medial cancellous bone for 30 mm distal to the femoral neck cut increases cement mantle thickness, the incidence of cement fractures.

COMMENT
Best results for femoral components allow for 2–5 mm proximal–medial thickness of cement mantle, less than 2 mm of proximal–medial cancellous bone thickness, a stem that fills more than half the distal part of the medullary canal and a stem in neutral orientation. Worst results for femoral components occur with a cement mantle thickness > 10 mm, a femur with more than 2 mm proximal–medial cancellous bone, a stem that fills half orless of the medullary canal and those in varus orientation.
EXAMINER
What would you inform your juniors about too litile or too much cement?
CANDIDATE
The femoral stems of hips that have a 2–5 mm thick cement mantle in the proximal medial region have a better outcome than stems implanted.
EXAMINER
What about the cement, how can this be improved?
CANDIDATE
improved interface adherence) to increase strain resistance, microfractures should lead to improved long-term results from cement fixation.
EXAMINER
What measures can betaken intra operatively to improve the quality of the cement mantle?
CANDIDATE
source p. 1323
Figure
Figurep. 1323

Figure 21.18 Anteroposterior radiograph showing the Exeter total hip replacement with radiolucent lines around the femoral component and fracture of the cement mantle.

1. Canal preparation

Use of correctly sized broaches that allow a mantle of adequate thickness.

hypotensive analgesia to reduce bleeding, avoidance of blood/cement occlusions.

2. Cement preparation

Occlusion of the canal using a distal plug, retrograde filling of the canal and cement gun pressurization of the cement column with a tight proximal seal are essential in achieving an interdigitating , uniform and homogeneous cement mantle.

EXAMINER
How can the cement mantle be optimized How can we obtain a high-quality cement mantle?
source p. 1324
CANDIDATE
Methods forcement fixation optimization include: Cement gun pressurization (enhances interdigitation) of the cement column with a tight proximal seal (femoral pressurizer). Porosity reduction (v acuum-mixing) which leads to reduced stress points in cement. Cement mantle thickness > 2 mm. Stem centralizer (reduces risk of stem malpositioning to decrease stress on the cement mantle).
EXAMINER
Have you heard of boneloc bone cement?
CANDIDATE
This is a bone cement that was withdrawn quite soon after introduction because of unacceptable revision rates with its use.
EXAMINER
What is the ideal cement mantle thickness?
CANDIDATE
I would aim for a cement mantle thickness greater than 2 mm as any less than this increases the risk of cement mantle fracture.
EXAMINER
Have you heard of the French paradox?
CANDIDATE
No, I am sorry I haven’t. Int heUK and USA, the first technique aims to produce a complete cement mantle of at least 2 mm in thickness and without ‘windows’. This has been called the French paradox in which implantation of a canal-filling femoral component in a line-to-line manner is associated with a thin cement mantle.21
source p. 1325

a canal-filling stem was supported mainly by cortical bone, subjected to low stresses.

COMMENT
Barracks grading of cement.22,23 Generations of cementing technique. Categories of loosening of cemented stems (Harris). Exeter vs Charnley stem design.
source p. 1326

Structured oral examination question 9#

source p. 1327

Wear in TKA

EXAMINER
What types of wear occur in TKA?
CANDIDATE
There are two main types of tibial component wear in TKA, (1) adhesive and abrasive wear and (2) fatigue damage (piting and delamination ).
EXAMINER
Fatigue w ear in tibial poly – what, why, where?
CANDIDATE
The appearance of fatigue w ear damage is primarily associated with cyclic compressive- tensile loading at the bearing surface, which generates subsurface tensile stresses that initiate and propagate cracks to form delamination and piting damage. Instances of fatigue w ear and fracture have repeatedly occurred over the history of UHMWPE use in TJR due to changes in molecular weight, fusion defects, crystallinity or cross-linking that can reduce the polymer’s resistance to crack initiation and growth. Fusion defects can exist along particle boundaries, acting as cr ack nucleation sites for fatigue w ear (delamination orpit ing) under cyclic sliding contact. This renders tibial PE inserts more susceptible to fatigue fracture or delamination w ear in the presence of high cyclic contact stresses or at sites of stress concentration, especially in a non-c onforming bearing surface.
source p. 1328

With melting of HCL PEther e is lowered crystallinity with improved wear characteristics but reduced mechanical properties.

Annealing avoids the reduction in the crystalline structure, but there is incomplete elimination of free radicals.

COMMENT
The JBJS review article ‘Osteolysis complicating total knee arthroplasty’24 provides a good framework for this viva topic.
Figure
Figurep. 1328

Figure 21.19 Tibial PE insert demonstrating fatigue failure. Fatigue failure is the formation of subsurface cracks in the polyethylene caused by cycles of loading and unloading of the joint, which then propagate and create particles that are shed into the joint space.

source p. 1329

Structured oral examination question 10#

source p. 1330

Osteolysis

The candidate is shown a picture of a THA with femoral osteolysis (Figure 21.20).

Figure 21.20
Figure 21.20Figure 21.20 Anteroposterior (AP) radiograph, pelvis, demonstrating femoral osteolysis.p. 1331
COMMENT
This is essentially a question on osteolysis and wear. Wear debris from prosthetic materials or bone cement is phagocytosed by macrophages causing release of various mediators.
EXAMINER
What size of PE material?
CANDIDATE
Studies have demonstrated that wear particles phagocytosed by macrophages elicited different responses depending on particle size Particles measuring < 5 μm in size generated a strong mononuclear macrophage response, whereas larger particles resulted in more multinucleated giant cells. The debris generated from prosthetic w ear triggers a cascade of macrophage cytokines, such as interleukin-1-beta (IL-1β) and tumour necrosis factor-alpha (TNF-α), among others, resulting in osteoclastic bone resorption and eventually leading to osteolysis. Material properties (HXLPE vs UHMWPE): higher percentages of small wear particles (0.1 –1 μm range) are produced during laboratory wear of HXLPE than conventional PE. The cellular mechanism of particle-induced osteolysis is that macrophages in the periprosthetic tissues phagocytose wear particles and become activated releasing an array of cytokines, leading to increased osteoclastic resorption of the adjacent bone and the production of the granulomatous tissue that fills the resorbed space.
source p. 1331

The majority of bone resorption occurs from osteoclasts recruited to sites of osteolysis and activated by the osteoclastogenic molecules.

The major pathway of osteoclastogenesis is the production of RANKL by osteoblastic stromal cells and binding of RANKL to its cognate receptor, RANK, on the surface of osteoclast precursors, stimulating these cells to differentiate into mature, active osteoclasts capable of resorbing bone.

Other cell types, including fibroblasts, osteocytes and activated T cells, also produce RANK Land are capable of stimulating osteoclast ogenesis.

Exposure of osteoblast-like cells toPE has been shown to induce changes in the rate of cell proliferation, to decrease alkaline phosphatase activity , and to increase the production ofos teoclastogenic mediators, such as PGE2, IL-6, GM-CSF, RANK Land nitric oxide.

EXAMINER
Types of wear and modes of wear?
COMMENT
See previous viva questions.
Figure
Figurep. 1331

Figure 21.20 Anteroposterior (AP) radiograph, pelvis, demonstrating femoral osteolysis.

source p. 1332

Structured oral examination question 11#

source p. 1333

Implant materials

COMMENT
Total knee arthroplasty components and the materials used. Discussions about the advantages of cobalt chrome versus stainless steel and then polyethylene manufacture and sterilization. The implant biomaterials used in total knee arthroplasty include: Stainless steel. Its use in TK Awas restricted because other metallic alloys such as Ti-based and Co–Cr-based alloys exhibited superior mechanical (yield strength) and corrosion properties. Newer implant stainless steel contains a high chromium, molybdenum and nitrogen content, making it stronger and resistant to local corrosion. There are several methods that can be used to modify Young’s elastic modulus of stainless steel.
EXAMINER
What does the 316L stand for in stainless steel?
CANDIDATE
The 3 stands for molybdenum (3%), 16 for nickel (16%) and L for low carbon (any stainless steel with less than 0.03% carbon).
source p. 1334

Cobalt chromium

Used as a bearing surface as it is very smooth and scratch resistant. It has a high Young’s modulus of elasticity and therefore risk of stress shielding.

Titanium

Young’s modulus of elasticity is closer to bone; therefore, it is more ductile, with good corrosion resistance, ability to integrate with bone, inert, biocompatible and extremely strong.

Expensive.

EXAMINER
Can you give me an example of a titanium alloy used?
CANDIDATE
Titanium alloy 6AL4V (titanium 89%, aluminium 6%, vanadium 4%, others 1%). Oxidized zirconium (OxZr): developed as an alternative bearing material for TJA. Higher-order thinking (HOT): puting this all together, cobalt–chromium alloys remain the predominant material (gold standard) used for TKA. There is a worry of increased PE tibial back side wear if titanium is used in place of cobalt–chromium. All poly tibia inserts were introduced to reduce wear and cost, but had poorer clinical results than conventional me tal tibia baseplates.
source p. 1335

Structured oral examination question 12#

source p. 1336

Wear and osteolysis

Acetabular cup. Explanted and worn.

What side is up? Divots on the other side may be from neck impingement.

How can you prevent wear? Implant factors/surgical factors.

Implant material, poly manufacturing process and direct compression moulding.

What about the head – we could use ceramics as less rough and better scratch profile.

Tell me about problems with wear – go through the whole RANK/RANKL discussion.

EXAMINER
What side is up?
CANDIDATE
There is eccentric PE wear in the cup superiorly (Figure 21.21).
Figure 21.21
Figure 21.21Figure 21.21 PE cup demonstrating superior eccentric wear.p. 1338
EXAMINER
What is the difference between PE creep and wear (Figure 21.22)?
Figure 21.22
Figure 21.22Figure 21.22 Difference between PE cup creep and wear.p. 1339
CANDIDATE
Creep is normal loading of the polyethylene cup and is superomedial.
EXAMINER
This worn bit here.
CANDIDATE
I pointed out divots on the outer side of the PE cup and said this could be due to neck impingement.
EXAMINER
How do you prevent impingement?
CANDIDATE
Impingement in THA is both implant- and surgeon-dependent. The implant design factors are those that influence the femoral head–neck ratio aswell as features of acetabular design.
EXAMINER
What factors affect the head–neck ratio?
CANDIDATE
Cam-type impingement can occur with use of a small head on a large circular taper or the use of a skirted femoral head.
source p. 1337
EXAMINER
What features increase acetabular impingement?
CANDIDATE
Features that increase acetabular impingement include the chamfer geometry of the rim of the polyethylene and the presence of an extended-rim (hooded) liner, particularly if the hood is incorrectly positioned in the hip.
EXAMINER
What do you mean by chamfer geometry?
CANDIDATE
The chamfer geometry is where the liner rim is sloped.
EXAMINER
How do you reduce the chances of impingement occurring?
CANDIDATE
Correct restoration of femoral offset and leg length.
EXAMINER
How do you ensure correct leg length and restoration of femoral offset?
CANDIDATE
It is essential to template the hip preoperatively and use a calliper-type pin device intra operatively to check leg length.
EXAMINER
How can you prevent wear?
COMMENT
This is best discussed in terms of patien t-related factors, implant-related factors and surgical factors (see above). Implant material – the poly manufacturing process that has the best wear profile is direct compression moulding into the shape of the desired product. Conventional PE used in hip arthroplasty was sterilized by gamma radiation in air , which offered the benefit of cross-linking but at the same time, this process produced free radicals that oxidized in air, leading to increased wear. High-dose gamma irradiation of polyethylene is not a sterilization process but a procedure to produce highly cross-linked PE.
source p. 1338
EXAMINER
Tell me about shelf oxidation.
CANDIDATE
Oxidative embritilemen t (characteristically identified as a subsurface white bandis atiribut ed to gamma sterilization in air.
EXAMINER
How does femoral head size affect wear?
CANDIDATE
With conventional UHMWPE the larger the femoral head the greater the volumetric wear. The smaller the head size the greater the amount of linear wear. Volumetric wear is proportional to the frictional torque of the THA. Therefore, an increase in femoral head size increases frictional torque and related volumetric wear.
Figure
Figurep. 1338

Figure 21.21 PE cup demonstrating superior eccentric wear.

source p. 1339
Figure
Figurep. 1339

Figure 21.22 Difference between PE cup creep and wear.

COMMENT
This is the basis of Charnley’s LFA.
EXAMINER
How can wear of a THA be measured?
CANDIDATE
1. Linear wear: the thickness of the acetabular cup decreases as it wears with use. Volumetric wear is related to linear wear by the equation Volumetric wear [mm3] = π × (radius of femoral head [mm])2 × linear wear [mm] As such, a larger diameter femoral head produces more volumetric wear for the same linear wear.
EXAMINER
Can you tell me about the scratch profile of ceramic compared to metal (Figure 21.23)?
Figure 21.23
Figure 21.23Figure 21.23 Scratch profile of metal and ceramic heads.p. 1340
COMMENT
Best to draw this out if allowed (Figure 21.24). Scratches of the femoral head can lead to an increased rate of PE wear. Additionally , ceramic has a more rounded surface profile with fewer sharp ridges than a metal surface, thus making it better suited for a bearing surface.
Figure 21.24
Figure 21.24Figure 21.24 Candidate drawing of scratch profile.p. 1340
source p. 1340

Ceramic is chemically inert. In the aqueous environment of the body, passive oxide films form on the surface of metal femoral heads. These consequences are avoided with ceramic heads.

Figure
Figurep. 1340

Figure 21.23 Scratch profile of metal and ceramic heads.

Figure
Figurep. 1340

Figure 21.24 Candidate drawing of scratch profile.

EXAMINER
What factors affect wear of a hard-on-soft bearing surface?
CANDIDATE
For the head (hard-bearing surface): Surface roughness. These areas increase PE wear. Hardness (scratch resistance, adhesive wear). Its modification by irradiation.
source p. 1341

References

1. Buzzword.

2. Buzzword again.

3. This applies in general to most of your answers. Unnecessary.

Although the examiners are more interested in the different types of mechanical wear, we would still mention it.

5. Analogies are helpful for the purpose of explanation or clarification.

6. Lee S-S, Purdue PE, Nam J-S. Inflammatory periprosthetic bone loss. In Inflammatory Diseases – Immunopathology, Clinical and Pharmacological Bases. IntechOpen; 2012. DOI: 10.5772/25558.

7. Banaszkiewicz PA. 2014, Springer. p. 85–87

8. Schmalzried T, Jasty M, and Harris WH. JBJS. 1992;74(6):849–863.

9. Quite common in practice viva sin underprepared candidates a few weeks away from siting the actual part 2 exam.

10. This may be the main thrust of the topic.

11. This is a gift.

12. Kim Y-H, Park JW, Kim JS, Lee JH. 2015;473(11):3588–3594.

13. Kindsfater KA, Pomeroy D, Clark CR, Gruen TA, Murphy J, Him denS. J Arthropl. 2015;30(8):1333–1338.

14. Lachiewicz PF, Soileau ES. A randomized trial. Clin Orthop Rel Res. 2016;474(1):88–95.

15. Rules need to be broken if needs be.

source p. 1342

16. Hood RW, Wright TM, Burstein AH. J Biomed MatRes Part A. 1983;17(5):829–842.

17. Doran J, YuS Smith D, Iorio R. 2015;28(05):382–389.

18. AhmedI, Salmon LJ, Waller A, Watanabe H, Roe JP, Pinczewski LA. 2016;98(1):58–64.

19. Dennis DA, Lynch CB. Optimizing the femoral component cement mantle in total hip arthroplasty. Orthopedics. 2005;28(8):S867–871.

20. Ebramzadeh E, Sarmiento A, McKellop HA, et al. J Bone Joint Surg Am. 1994;76:77–87.

21. El Masri F, Kerboull M, Kerboull L, Courpied JP, Hamadouche M.

22. Barrack RL, Mulroy R, Harris WH. A 12-year radiographic review. 1992;74(3):385–389.

23. Banaszkiewicz PA. In Classic Papers in Orthopaedics. 2014, Springer. p. 31–34.

24. Gilbert TJ, Anoushiravani AA, Sayeed Z, Chambers MC, El-Othmani MM, Saleh KJ. JBJS Rev.

figure